Specification Information
Transportation and
storage conditions
Operating conditions
Standards for
electrical equipment
Co m bi n a ti o n W ar m er
Op e r at or ' s Man u a l
▪
Model
Intended use
Net: 105 lb (48 kg)
Weight
Ship: 185 lb (84 kg)
TriMode cavity: 24 1-liter bottles
Storage cavity
or 20 1-liter bags, or
capacity
9–12 blankets (4.75 ft
Temperature
range
Clearance
requirements
▪
Ambient temperature range of -40°F to +159°F (-40°C to +70°C).
▪
Relative humidity range of 10% to 95%, non-condensing.
▪
Atmospheric pressure range of 7.25 psi to 15.37 psi (50kPa to 106kPa).
▪
The warmer must acclimate to the room temperature it will be placed in—24
hours is recommended.
▪
The recommended environmental temperature range is 60°F to 90°F
(15°C to 32°C).
▪
The recommended relative humidity is above 20%, non-condensing.
▪
Medical equipment listed by Underwriters Laboratories with respect to
electrical shock, fire, and mechanical hazards only, in accordance with UL
61010-1 and CAN/CSA C22.2 No. 61010-1.
▪
Grounding reliability can only be achieved when the appliance is
connected to an equivalent receptacle marked "Hospital Grade."
▪
IP-XO ordinary
MN -4 7 42 4
Re v 1
▪
▪
▪
P-2420
18" Depth
Warming blankets, irrigation or injection fluids
3
)
Irrigation fluids (IRR): 90°F to 150°F (32°C to 66°C)
Injection fluids (INJ): 90°F to 104°F (32°C to 40°C)
Blanket (BLNKT): 90°F to 160°F (32°C to 71°C)
0.5" (12.7mm) from rear
0.5" (12.7mm) from top
0.5" (12.7mm) from sides
4 / 2 1
S
PECIFICATIONS
P-2430
24" Depth
Net: 135 lb (61 kg)
Ship: 215 lb (98 kg)
TriMode cavity: 30 1-liter bottles
+ 36 1/2 liter-bottles, 48 1-liter
bags, or 10–14 blankets (5.5 ft
Protective Earth
Ground Symbol
3
)
E471516
13