Pedigo P-2150 Operator's Manual page 48

Combination blanket/fluid warmer
Table of Contents

Advertisement

T
ROUBLESHOOTING
Code
Refers to
E-11
Cavity sensor
ES11
Pad sensor 1
ES21
Pad sensor 2
ES31
Pad sensor 3
ES41
Pad sensor 4
ES51
Pad sensor 5
ES61
Pad sensor 6
ES71
Pad sensor 7
E-30
Under
temperature
*E-31
Cavity sensor
P131
Pad sensor 1
P231
Pad sensor 2
P331
Pad sensor 3
P431
Pad sensor 4
P531
Pad sensor 5
P631
Pad sensor 6
P731
Pad sensor 7
*E-33
Cavity sensor
*E-50
Analog to Digital
Converter Error
48
Action required
Sensor is open. Software disengages heating pads. Acknowledge error by
pressing the standby button. If error persists, a qualified service technician
should test the sensor.
◾Test the sensor. Disconnect the sensor from the warmer control PCB
and use an Ohm meter to measure the resistance of the sensor.
Sensor PR-37140 should measure 10 KOhm ± 1.5 KOhm at 25°C (77°F).
Sensor SN-33541 should measure 100 Ohm ± 10 Ohm at 25°C (77°F).
Replace sensor if measurement is outside of tolerance.
◾Check the wires for integrity. Check for proper and secure connections
at the control and terminal block. If necessary, re-secure the faulty
connections.
◾Contact service if error persists.
The blanket cavity temperature is lower than the set temperature for 90
minutes or longer.
◾Make sure the door is closed.
◾If the cavity is overloaded, redistribute the inventory. Do not exceed the
height of the insert.
◾Test the sensor. Disconnect the sensor from the warmer control PCB
and use an Ohm meter to measure the resistance of the sensor.
Sensor PR-37140 should measure 10 KOhm ± 1.5 KOhm at 25°C (77°F).
Sensor SN-33541 should measure 100 Ohm ± 10 Ohm at 25°C (77°F).
Replace sensor if measurement is outside of tolerance.
◾The sensor reading is above the temperature set-point. Blanket warmer
triggers at 15° over set-point. Fluid warmer triggers at 5° over set-point.
◾The difference between the room temperature and the fluid set-point
temperature must be greater than 11°C (20°F).
◾Contact service if error persists.
Heater pad over-temp error.
◾Software disengages heating pads.
◾Acknowledge error by pressing the standby button.
◾Allow the warmer to cool.
◾Contact service if the error persists.
◾Sensor reading is above maximum allowable temperature set-point and
over temp value. Blanket warmers trigger at 82°C (180°F). Fluid warmers
trigger at 71°C (160°F).
◾Contact service if error persists.
◾Remove inventory, discard if necessary, and allow the warmer to cool
down.
◾If error persists after cool down and reset, the control assembly should
be replaced by a qualified service technician. Contact service.
Co m b i n at i on B l an k e t/ F l u id W a rm e r
Op er a to r ' s Ma n u al
M N- 39 9 3 0
Rev 2
4/ 2 0