Pedigo P-2150 Operator's Manual page 50

Combination blanket/fluid warmer
Table of Contents

Advertisement

T
ROUBLESHOOTING
Code
Refers to
*EFAn
Fan or Fan Sensor
Failure
Red iv fluid
Red LED is
temperature
illuminated within
status
iv storage shelf
indicator
50
Action required
◾Check to make sure the fan sensor wires did not disconnect from the fan
or the control board.
◾If the error persists after checking the wires, replace the fan.
◾Set the power switch to the OFF position.
◾Remove any iv bags from inside the warmer.
◾Set the power switch to the ON position.
◾Visually check to see if red LED remains illuminated.
◾Contact service if error persists.
*
All non-critical codes can be cleared using the standby button. Critical errors (marked with a *) can
only be cleared by setting the power switch at the rear of the warmer to the OFF position and
allowing the warmer to cool.
Co m b i n at i on B l an k e t/ F l u id W a rm e r
Op er a to r ' s Ma n u al
M N- 39 9 3 0
Rev 2
4/ 2 0