Manual EN COMTRAXX® CP9xx – Control Panel Remote alarm indicator and operator panel for medical locations and other areas Software version V4.9.x CP9xx_D00349_10_M_XXEN/07.2024...
Signs and symbols............................5 Service and Support............................5 Training courses and seminars........................6 Delivery conditions............................6 Inspection, transport and storage......................6 Warranty and liability.............................6 Disposal of Bender devices..........................7 1.10 Safety..................................7 Intended use........................8 Product description......................9 Scope of delivery............................. 9 Device features..............................9 Possible applications............................10...
Page 4
Table of contents Commissioning of the device........................23 4.7.1 BMS interface..............................24 4.7.2 Address configuration and termination....................24 4.7.3 Browser configuration..........................24 4.7.4 Software products used..........................24 COMTRAXX® user interface........................25 Modbus TCP server......................26 Troubleshooting......................27 Malfunctions..............................27 6.1.1 What should be checked?..........................27 6.1.2 Frequently asked questions........................
Protect from dust Temperature range Recycling RoHS directives Service and Support Information and contact details about customer service, repair service or field service for Bender devices are available on the following website: Fast assistance | Bender GmbH & Co. KG. CP9xx_D00349_10_M_XXEN/07.2024...
Regular face-to-face or online seminars for customers and other interested parties: www.bender.de > know-how > seminars. Delivery conditions The conditions of sale and delivery set out by Bender GmbH & Co. KG apply. These can be obtained in printed or electronic format. The following applies to software products: "Software clause in respect of the licensing of standard software as part of deliveries,...
COMTRAXX® CP9xx – Control Panel Disposal of Bender devices Abide by the national regulations and laws governing the disposal of this device. For more information on the disposal of Bender devices, refer to www.bender.de > service & support. 1.10 Safety If the device is used outside the Federal Republic of Germany, the applicable local standards and regulations must be complied with.
These include, for example: • All Bender devices with BMS bus or BCOM interface • Bender devices (PEM, energy meters,…) with Modbus RTU or Modbus TCP interface • Other devices with Modbus RTU or Modbus TCP interface In addition, the data is available via Modbus TCP, Modbus RTU, SNMP, MQTT and PROFINET protocols. This allows coupling to a higher-level building control system as well as visualisation and evaluation using standard web browsers.
Included within the scope of delivery • A CP9xx alarm indicator and operator panel • A printed quick-start guide • Safety instructions for Bender products • The manuals "COMTRAXX® CP9xx" and "BCOM" are available as PDF files for download at https:// www.bender.de/en/service-support/download-area/ For CP915 and CP924 additionally •...
Product description Possible applications Monitoring, operation and display of • Medical IT systems • Supply systems for medical gases • HVAC systems • Room lighting • Operating theatre lights • Special power supply systems (BSV or UPS) • Other systems from different manufacturers Configuration, diagnosis, service Each panel can be individually manufactured and tailored to the requirements of the user.
CP9xx communicate with the devices and systems assigned via various interfaces: • Internal BMS bus (RS-485) for Bender systems such as EDS46…/49…, RCMS46…/49… and MEDICS®. CP9xx can be operated as a master or as a slave. When operated as a master, requests are answered more quickly.
Product description 3.8.2 Process image The CP9xx alarm indicator and operator panel combines the information from the different interfaces and makes it available for operation and visualisation via the web user interface of a PC. It acts as a central user interface. In order to be able to identify them, one individual address is available for each device on this interface.
COMTRAXX® CP9xx – Control Panel Mounting, wiring and commissioning Only qualified personnel are permitted to carry out the work necessary to install, commission and run a device or system. CAUTION Protective earth The device must be earthed. Without connection of the protective earth, the device function is not guaranteed.
COMTRAXX® CP9xx – Control Panel Flush mounting enclosure CP907, dimensions in mm Installation dimensions flush-mounting enclosure = enclosure dimensions + 3 mm 4.2.2 Surface-mounting enclosure CP907 For surface mounting, the flush-mounting enclosure is mounted in the optionally available surface-mounting enclosure (B95061915). Enclosure for surface mounting CP907, dimensions in mm Mounting procedure Assemble the surface-mounting enclosure (fit hinges and bracket).
Mounting, wiring and commissioning Mounting CP915/CP924 4.3.1 Flush-mounting enclosure CP915/CP924 Mounting For UL applications In case of flush-mounting a suitable and approved housing must be used. Dimensions flush-mounting enclosure Flush-mounting enclosure CP915/CP924, dimensions in mm Installation dimensions flush-mounting enclosure = enclosure dimensions + 3 mm 4.3.2 Removing the CP915/924 front plate Removing the front plate...
COMTRAXX® CP9xx – Control Panel Connection of the device The CP9xx is integrated into existing LAN structures, but can also be operated via a single PC. Configuration of computer networks If you are familiar with the configuration of computer networks, you can carry out the connection of the device yourself.
Page 18
Mounting, wiring and commissioning Connection CP907 CP915/CP924 Plug connector digital inputs Plug connector to energy storage board Voltage supply A1/+, A2/–, PE* Ethernet (RJ45/CAT5); HTTP, Modbus TCP, BCOM with PoE without PoE X1 plug connector for Modbus RTU, BMS bus Termination of Modbus RTU and BMS bus USB ports (for touch sensor) not included...
Page 19
COMTRAXX® CP9xx – Control Panel Establish connection with PC and BCOM: Connect the CP9xx device to the PC network using an Ethernet cable (4). Link digital inputs. See chapter “Digital inputs”, page 22. Connect the control relay (11): Connection relay N/C operation contacts 11-12 (the alarm relay is energised during normal operation).
Page 20
Mounting, wiring and commissioning The CP907 can also be supplied via Power-over-Ethernet (PoE). The PoE switch must be earthed. For further details, see technical data. Attach the front plate to the built-in flush-mounting enclosure. CP907 is mounted. The following steps apply to CP915/924 only: Connect the front panel to the control board and the power supply: Connect a USB socket (7) to the touch sensor connector on the front panel.
Page 21
COMTRAXX® CP9xx – Control Panel The dual power supply unit supplies the CP915 or CP924 with 24 V and the display with 12 V. 13. CP915: Attach the front plate to the built-in flush-mounting enclosure. 14. CP924: Hold the display unit from above in front of the flush-mounting enclosure. Place the two mounting hooks behind the upper snap locks of the flush-mounting enclosure (see red markings).
Mounting, wiring and commissioning Digital inputs CP9xx-devices feature 12 configurable digital inputs. Settings are made via the COMTRAXX® user interface in a browser. Function The following functions can be assigned to the digital inputs: Digital input without function TEST Self test of the device RESET Reset fault and alarm messages Menu >...
COMTRAXX® CP9xx – Control Panel Relay Connection relay N/C operation contacts 11-12 (the alarm relay is energised during normal operation). N/O operation contacts 11-14 (the alarm relay is de-energised during normal operation). Commissioning of the device Switch on the supply voltage: After switching on, the device performs a start routine.
After this, the commissioning must be continued via the COMTRAXX® user interface. 4.7.1 BMS interface The majority of Bender devices communicate via the internal BMS bus. CP9xx can be operated as a master or as a slave. CP9xx is to be operated as a master if: •...
COMTRAXX® CP9xx – Control Panel COMTRAXX® user interface The device has a web user interface for setting and operation. How to start the web user interface: • Open an Internet browser from any network device. • Enter the address of the CP9xx device in the address line of the browser. It is possible to connect the CP9xx directly to a computer/laptop.
Modbus TCP server Modbus TCP server The Modbus TCP server supports the following function codes: • Function code 0x03 (Read Holding Registers) • Function code 0x04 (Read Input Registers) • Function code 0x10 (Preset Multiple Registers) The Modbus TCP server generates a function-related response to requests and sends it back to the Modbus TCP client.
Document the new settings as a PDF file. Use the backup function to save all settings of the device (see Chapter: “Device features”, page 9). Frequently asked questions on the Internet FAQs on many Bender devices can be found at: www.bender.de/en > service-support > fast-assistance Device operation, Maintenance, Cleaning Device operation The device can be operated with latex, vinyl and nitrile gloves without impairing functionality.
Technical data Technical data Factory settings Factory settings communication addresses Parameter Factory setting IP address IP address for 1:1 ETH conn. 169.254.0.1 Net mask 255.255.0.0 Standard gateway 192.168.0.1 DHCP Timeout for address assignment 30 s BMS address BMS protocol BMS i BCOM system name SYSTEM Subsystem address...
Page 29
COMTRAXX® CP9xx – Control Panel Supply CP907 via plug-in terminal (A1/+; A2/-) Nominal voltage DC 24 V SELV/PELV Nominal voltage tolerance ±20 % Typical power consumption at DC 24 V < 15 W Maximum cable length when supplied via B95061210 (24-V DC power supply unit 1.75 A): 0.28 mm²...
Page 30
Technical data Displays, memory Display CP907/Resolution 7“ TFT-Touch Display/800 x 480 Display CP915/Resolution 15,6“ TFT-Touch Display/1280 x 720 Display CP924/Resolution 24“ TFT-Touch Display/1280 x 720 or 1920 x1080 E-mail configuration and device failure monitoring max. 250 entries Individual texts unlimited number of texts with 100 characters each Number of data points for "third-party devices"...
Page 31
Modbus TCP Interface/protocol Ethernet/Modbus TCP Cable length < 100 m Operating mode client for Bender Modbus TCP devices and "third-party devices" Operating mode Server for access to process image and for Modbus control commands Parallel data access from different clients max.
Page 32
Technical data Modbus RTU Terminating resistor 120 (0.25 W), can be connected internally (see plug-in terminal) Supported Modbus RTU slave addresses 2…247 PROFINET Interface/protocol Ethernet/PROFINET Operating mode slave (IO device) SNMP Interface/protocol Ethernet/SNMP Versions 1, 2c, 3 Supported devices query of all devices (channels) possible Trap support MQTT Interface/protocol...
Page 33
COMTRAXX® CP9xx – Control Panel 48862 BCOM (UDP) Digital inputs (1…12) Number Galvanic separation Maximum cable length < 1000 m Operating mode selectable for each input: active-high or active-low Factory setting active-high Voltage range (high) AC/DC 10…30 V Voltage range (low) AC/DC 0…2 V Max.
Page 34
Technical data Audio Line IN not used Line OUT Output to a STEREO playback device via 3.5 mm jack plug Cable length < 3 m Device connections Terminal block (L1; N; PE) (for CP915 and CP924 only) Conductor sizes AWG 20…12 Stripping length 10…11 mm rigid/flexible...
Page 35
COMTRAXX® CP9xx – Control Panel Environment/EMC IEC 61326-1 Operating temperature CP907 -10…+55 °C CP907 for UL applications -10…+50 °C CP915 -5…+40 °C CP924 -5…+40 °C Operating altitude ≤ 2000 m AMSL Rel. humidity ≤ 98 % at 25 °C Classification of climatic conditions acc. to IEC 60721 Stationary use (IEC 60721-3-3) 3K22 Transport (IEC 60721-3-2)
Technical data Dimensions CP907 (B x H x T) 226 x 144 x 78 mm CP915 (B x H x T) 505 x 350 x 95 mm CP924 (B x H x T) 654 x 441 x 100 mm Weight CP907 <...
COMTRAXX® CP9xx – Control Panel Individual components Device series Type Art. No. CP907 Flush-mounting enclosure B95100140 Display unit, white B95061112 CP915 Display unit, grey B95061113 Display unit, white B95061115 CP924 Display unit, grey B95061116 Accessories Device series Type Art. No. CP907 Surface-mounting enclosure B95061915...
Page 38
Technical data CP9xx_D00349_10_M_XXEN/07.2024...
Page 39
COMTRAXX® CP9xx – Control Panel CP9xx_D00349_10_M_XXEN/07.2024...