Maintenance mode

The user can change the configuration of the instrument by accessing the Maintenance mode.
1. Press
2. Wait a few seconds and press the
System menu.
From the Maintenance System menu you can access the System Settings, the
System Info, the Configuration and the Calibration date. You are also presented to
a selection of adjustment options.
The editable possibilities in the System Settings and the System Info are the same
as described in section 5.1.1.
Configuration
If the system is locked on a specific main frequency it is possible to change this
frequency.
The alarm list will be displayed with the following text:
Wrong mains frequency
Note...
In order to change the main frequency a pin code is needed. Please contact your
local distributor to obtain your special pin code.
1. Select the Configuration mode.
Type in the pin code obtained from the distributor using the
2.
(Down) keys. The pin code must consist of 4 digits.
129375 06
(Enter) while switching on the calibrator using the ON/OFF switch.
(System) key to access the Maintenance
2019-12-06
(Up) and
44