battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.

Flat battery - replacement

When the batteries are flat, the battery symbol AM will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment 6 at the back of the device.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in Section
«2. Using the device for the first time».
The measurements stored in the memory are deleted when
the batteries are removed from the battery compartment
(e.g. when replacing batteries).

Which batteries and which procedure?

Use 4 new, long-life 1.5 V, size AAA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.

8. Error messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description
«ERR 1»
Signal too weak
AR
«ERR 2»
Error signal
AQ-B
Microlife BP B1 Classic
Potential cause and remedy
The pulse signals on the cuff
are too weak. Re-position the
cuff and repeat the measure-
ment.*
During the measurement,
error signals were detected
by the cuff, caused for
instance by movement or
muscle tension. Repeat the
measurement, keeping your
arm still.
Error
Description
«ERR 3»
Abnormal cuff
pressure
AQ-C
«ERR 5» Abnormal result
«HI»
Pulse or cuff
pressure too high
«LO»
Pulse too low
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.

9. Safety, care, accuracy test and disposal

m
Safety and protection
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical specifications» section.
Potential cause and remedy
An adequate pressure cannot
be generated in the cuff. A
leak may have occurred.
Check that the cuff is
correctly connected and is not
too loose. Replace the
batteries if necessary. Repeat
the measurement.
The measuring signals are
inaccurate and no result can
therefore be displayed. Read
through the checklist for
performing reliable measure-
ments and then repeat the
measurement.*
The pressure in the cuff is too
high (over 299 mmHg) OR
the pulse is too high (over 200
beats per minute). Relax for 5
minutes and repeat the
measurement.*
The pulse is too low (less than
40 beats per minute). Repeat
the measurement.*
5
EN