50 | DX-D Software Console | Problem solving
Information message Description
Max mAs
Min ms
Max ms
Min ms & Min mA
Max ms & Max mA
Max Power
Space Charge
Max Energy (60kJ)
Instant Power
0189D EN 20180329 1421
Maximum mAs = 500 mAs (regulatory limit for
AEC)
Minimum exposure time (generator limit)
Maximum exposure time (generator limit)
Minimum exposure time and minimum mA
(generator limit)
Maximum exposure time and maximum mA
(generator limit)
Maximum power (generator limit)
Filament emission limit for a combination of kVp
and mA in the selected focal spot.
The maximum energy cannot exceed 60 kJ (kVp x
mAs = 60 kJ) (regulatory limit for AEC).
Instantaneous power limit of the X-ray tube
depending on the anode temperature, the selected
exposure time and the selected focal spot.