Two Point Mode (2P) - AGFA DX-D User Manual

Table of Contents

Advertisement

42 | DX-D Software Console | Operating the DX-D Software Console

Two Point Mode (2P)

The value of kVp, mAs, max ms, the setting of focal spot and X-ray tube load
can be adjusted.
The value of mA and ms are adjusted automatically to keep the mAs value
constant, within the boundaries of generator or X-ray tube limitations.
The default value for mAs is 20 mAs.
The setting of density, sensitivity and patient size is not available.
By selecting one of the AEC field buttons, the one point mode is activated.
By adjusting the value of mA or ms, the three point mode is activated.
After exposure all values reflect the settings actually used by the generator.
Figure 19: 2P working mode
Related Links
Radiographic Parameters
0189D EN 20180329 1421
on page 33

Advertisement

Table of Contents
loading

Table of Contents