38 | DX-D Software Console | Operating the DX-D Software Console
Figure 14: Sensitivity
Density
These buttons are used to adjust the AEC cut-off dose (and patient entrance
dose accordingly).
Figure 15: Density
Density can be increased and decreased in a range of -4 to +4. Each step gives
a difference of 25% in dose (25% is the default setting). When disabled, the
density range number appears in black.
Table 9: Density scale variation over reference dose (0)
-4
-3
-2
-1
0
+1
+2
+3
+4
Patient Size
The size of the patient is classified in five categories: Extra Small, Small,
Medium, Large and Extra Large.
Touch the UP or DOWN arrows to select the desired patient size.
0189D EN 20180329 1421
Density
rd x 0.41
rd x 0.51
rd x 0.64
rd x 0.80
Reference dose (rd)
rd x 1.25
rd x 1.56
rd x 1.95
rd x 2.44