Function; Programs; Maximum Pressure (Caring Mode); Panel Lock - Care of Sweden CuroCell A4 Cirrus Instructions For Use Manual

Table of Contents

Advertisement

4.1

Function

The mattress system independently and without
manual adjustment controls the inner pressure
of the mattress according to the weight, height,
and position of the user. No manual action needs
to be performed to adjust the inner pressure of
the mattress. This function works as follows:
1. At start-up, automatic setting of the inner
pressure of the mattress is always carried out
according to the weight and height of the user.
2. If the user moves noticeably or changes their
position, the system will independently control
the inner pressure of the mattress.
3. The system performs an automatic setting at
fixed intervals even if no significant changes have
occurred.
After automatic setting of the mattress inner
pressure, the system returns to the previously
selected program. At start-up, Pulsating Mode is
always pre-set.
4.2

Programs

There are three programs to choose from:
CLP means that the air pressure in all of the air
cells is the same in the whole mattress.
Alternating mode means that the air pressure in
the air cells are different and alternates regularly
after chosen cycle periods.
Pulsating mode combinates CLP with the
alternating function.
Choose program by pushing the button for the
program. We recommend the Pulsating mode
which is also the preset mode.
1. Constant low pressure mode (CLP)
No cycle period is needed.
2. Alternating mode. The cycle period
can be changed according to the user
needs and requirements. Choose
between 10, 15, 20 or 25 minutes. The longer
the cycle period, the slower the alternations.
We recommend a basic setting of 10 minutes.
3. Pulsating mode. The cycle period can
be changed according to the user needs
and requirements. Choose between 10,
15, 20 or 25 minutes. The longer the cycle
period, the slower the alternations. We
recommend a basic setting of 10 minutes.
4.3

Maximum pressure (caring mode)

With this function, the entire mattress
is inflated and provides firmed
support. This function reverts to the
previous setting after 20 minutes. The
function should be used when caring the user,
shifting the user's position or moving the user in
or out of bed.
4.4

Panel lock

Press the Panel Lock button to lock or
unlock the control panel. The button
indicates when the panel has been
locked. The screen locks automatically if left
untouched for five minutes. This is to prevent
the settings being changed accidentally.
To unlock, press the button for 2 seconds.
4.5

Comfort settings

comfort requirements. This increase is made
based on the automatic setting in 5.1.
Note!
• When only parts of the mattress are under
load, for example, in the case of amputees
– it may be necessary to raise the setting using
the comfort settings.
The selected setting is shown by a green light.
8
CuroCell® A4 Cirrus
The pressure can be
increased in two steps
depending on the user's

Hide quick links: