Rear panel openings:
Fig.9: A sealable access port [1] allows cables, hoses
or additional sensors to be routed into the work space
of the device. A pressure compensation opening with
insert [2] at the rear panel of the device allows a com-
pensation between the pressures in the work space
and in the operating room.
NOTE!
Operating conditions:
When accessories are to be operated in the
work space of the CO
ent condition requirements must be obser-
ved (see table below). The energy
introduced into the work space affects the
beginning of the temperature control
range. When additional heating sources are
introduced into the work space, condensa-
tion (e.g. at the glass door) may occur.
Introduced
Control range of the temperature
energy
General
0 W
RT + 3 °C
5 W
RT + 6.5 °C
10 W
RT + 9.5 °C
15 W
RT + 13 °C
20 W
RT + 16 °C
*RT = Ambient temperature
incubator, the ambi-
2
Example:
RT* = 21 °C
24 °C
27.5 °C
30.5 °C
34 °C
37 °C
24

Fig.9: Rear panel openings