TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 INSTALLATION SECTION 3.0 CONTROL OVERVIEW SECTION 4.0 OPERATION SECTION 5.0 MAINTENANCE SECTION 6.0 TROUBLESHOOTING SECTION 7.0 PARTS LIST UNIT SPECIFICATIONS SCHEMATICS This unit is a general purpose Biochemical Oxygen Demand (BOD) incubator for professional, industrial or educational use where the preparation or testing of materials is done at approximately atmospheric pressure and no flammable, volatile or combustible materials are being heated.
Section RECEIVING AND INSPECTION Your satisfaction and safety require a complete understanding of this unit. Read the instructions thoroughly and be sure all operators are given adequate training before attempting to put the unit in service. NOTE: This equipment must be used only for its intended application;...
Section INSTALLATION Local city, county or other ordinances may govern the use of this equipment. If you have any questions about local requirements, please contact the appropriate local agency. Installation may be performed by the end user. Under normal circumstances this unit is intended for use indoors, at room temperatures between 5 and 40C, at no greater than 80% Relative Humidity (at 25C) and with a supply voltage that does not vary by more than 10%.
Section CONTROL PANEL OVERVIEW Power Switch: The main power I/O (on/off) switch controls all power to the unit and must be in the I/ON position before any systems are operational. Main Temperature Control: This control is marked SET TEMPERATURE and consists of the digital display and UP/DOWN arrow pads for inputting set point temperatures and calibration.
Section OPERATION Check power supply against unit serial plate. They must match. Plug service cord into the grounded electrical outlet. Push the power switch to the I/On position, and turn the Overtemperature Thermostat to its maximum position, clockwise using a coin or flat edged tool. Set Main Temperature Controller: Enter desired set point temperature.
Page 7
Load the chamber with sample bottles. See Figure 2 for bottle placement. Figure 2...
Section MAINTENANCE Note: Prior to performing maintenance and/or service on this incubator, disconnect the unit from the power supply. Cleaning: Cleaning and decontamination are recommended on a regular basis. To prepare the unit for cleaning, disconnect the power cord and remove all interior parts, such as shelves, trays, etc.
Section TROUBLESHOOTING FOR PERSONAL SAFETY, ALWAYS DISCONNECT THE POWER BEFORE SERVICING. In the event the incubator does not operate properly, check the following before calling for service. Always make a visual inspection of the incubator and control console when troubleshooting to find loose or disconnected wires which may be the source of trouble.
Page 10
2/ Be sure that fans are circulating air: in the chamber, and over the compressor 3/ Assure there is ample room around the unit as described in Installation. 4/ Compare ambient specifications to Unit specifications. If compressor isn’t running: 1/ Confirm that compressor cooling fan motor is operable a- Dirty coil or poor circulation b- Coil next to heat source c- Ambient temperature too high...
UNIT SPECIFICATIONS Weight Shipping 2005 130 lbs. 108 lbs. Exterior WxDxH Interior WxDxH Dimensions (in.) (in.) 2005 20.75 x 21.25 x 38.75 15 x 11 x 25 Capacity 3 cubic feet Temperature Range Uniformity 2005 0 to 45C +0.5C @ 20C...
Page 13
WIRE DIAGRAM 110V (9851370) 120VAC POWER CORD AND PLUG 1100505 CIRCUIT BREAKER 15A GREEN LIGHTED POWER SWITCH 7850570 OUTLETS 6100525 4880545 INTERNAL FAN SENSOR TEMPERATURE NEUTRAL CONTROL 1750967 GROUND 4650553 SAFETY ACTIVATED LOAD 4650554 HEATING ACTIVATED 45 1750715 155°F/68°C 1750862 9570863 ELEMENT ASSY 2350508 ELEMENT COIL THE COMPRESSOR IS PART...
Page 14
220V (9851371) POWER CORD 1800537 2800502 EMI FILTER FUSE FUSE GREEN LIGHTED POWER SWITCH 7850570 OUTLETS 6100526 TEMPERATURE NEUTRAL CONTROL 1750968 GROUND LOAD SAFETY ACTIVATED 4650553 HEATING ACTIVATED 4650554 1750715 155°F/68°C ½ 1750813 ½ 1750813 9570864 ELEMENT ASSY 2350508 ELEMENT COIL 8350515 TRANSFORMER 240/120VAC 750VA THE COMPRESSOR IS PART...
Page 15