TA Instruments TGA 2950 Operator's Manual page 195

Thermogravimetric analyzer
Table of Contents

Advertisement

TM
Hi-Res
Option
of undecomposed sample material which now
accelerates its rate of decomposition. As in the
case of too high a ramp rate, the increasing
temperature causes the remaining sample to
quickly decompose which raises the rate of
weight loss (%/minute) to a high level and either
triggers another isothermal hold period or shows
up as a backside shoulder on the weight loss
curve.
An example of using stepwise isothermal
TM
heating is shown in the following "Hi-Res
TGA Examples" section.