Updating the unit software

The unit software is updated using a PC program called SWUP. Before updating the
unit software, the update program must first be downloaded from the SCHILLER
server to a PC.
The procedure to update the unit software is a follows:
• The operating system of the PC must be Win NT4, Win 98, Win 2000 or Win XP.
• Do not power off during update.
RS-232 cable assembly.
1.
Access the SCHILLER extranet (details from the service department) and install
the software on your pc (Install_AT10plus.exe).
2.
Connect the RS-232 cable assembly between the PC and the RS-232 on the side
of the AT-10 plus.
3.
In the service screen click the Program Update icon (to enter receive mode).
4.
Start the AT10 plus SWUP program from the computer Start menu.
5.
Click the config icon (1).
6.
Define pc settings:
– Select the pc com port (2) to which the pc cable is connected.
– Select baudrate = auto.
7.
Click the update AT-10plus icon (3).
8.
The bargraph shows the progress of the download. After downloading, the device
will restart automatically.