Ecg Amplifier And Patient Cable Test; External Printer Test; Communication (Rs-232) Test - Schiller AT-10 plus Service Handbook

12-channel ecg device
Hide thumbs Also See for AT-10 plus:
Table of Contents

Advertisement

AT-10 plus
Service
Character
Colour
Printer Key
ECG
NIBP
Spiro
HW-
Lead Test
(mV)
RA
5
LA
5
C1
5
C2
5
C3
5
C4
5
C5
5
C6
5
C7
497
C8
495
C9
498
C10
498
LOFF
0F00
Service Handbook
5.3.8

ECG amplifier and patient cable test

1.
With a patient simulator connected, press the ECG key
Battery
Softwar
screen is displayed (this screen is also displayed in the service screen (see page
I/O +
20).
2.
Disconnect ECG patient simulator. Check the following:
– device beeps 4 times
– all lead designations highlighted
– the mV reading for all leads is -350 to -550 mV
3.
Connect ECG simulator and setup HR to 60 b/min, no arrhytmias.
4.
Check the following:
– all leads stops blinking
– the mV reading for all leads is -15 ...+15 mV
• When a standard 10-lead cable is connected, check RA, LA, and C1 to C6 only.
Additionally check C7, C8, and C9 when a 13 lead patient cable is used.
• Additionally check C7, C8, C9 and C10, when a 14 lead patient cable is used.
• LOFF is a register for factory use only (for lead confirmation).
5.3.9

External printer test

Only with the Comms Module. At the time of print this was not available
5.3.10

Communication (RS-232) test

Only with the Comms Module. At the time of print this was not available.
Functional Checks
Functional checks and tests
. The electrode
5
5.3