CARDlOVlT AT-6
Service Information
1.
EQUIPMENT NEEDED
Oscilloscope
Analog Meter
3 pieces of wire (approx.
6 )
1
short (low capacitance) tweaker
1
low capacitance screwdriver
soldering iron
1
small Phillips screwdriver
1
small standard screwdriver
2.
OPENING UP THE UNIT
1.
Pull out and remove the white contrast control knob from the right-hand side of the unit.
2.
Turn the unit over and remove the four Phillips screws in each corner of the unit.
You
will also need
to remove the two standard screws located in the centre, along the side, if they are present.
3.
Using both hands, hold the top and bottom cases together and turn the unit upright.
BE
CAREFUL
!
4.
Gently lift up on the front side of the top cover removing the cable connections as they appear.
Refer to Figure
1
for connector locations. This will expose the inside of the unit.
8-1