Battery disposal

Danger of explosion. Battery may not be burned or disposed of domestic refuse.
Danger of acid burns! Do not open the battery.
The battery is to be disposed of in municipally approved areas or sent back to
SCHILLER AG.

Changing the fuse and mains voltage

Before the fuse and mains voltage are changed, the device must be disconnected
from the mains and the mains plug removed for the wall socket.
The fuse may only be replaced by the fuse type the table below.

Fuse types

Voltage range
220 - 240 VAC
100 - 115 VAC

Changing a fuse

1.
Disconnect the device from the mains and remove the mains plug.
2.
Loosen the fuse inset using a screwdriver and remove it.
3.
Replace existing fuses with the same type. See table
4.
Re-insert the fuse inset.

Changing the mains voltage

1.
Disconnect the device from the mains and remove the mains
2.
Loosen the fuse using a screwdriver and remove it.
3.
Remove the grey inset, turn it by 180° and re-insert it.
4.
Check the voltage indication in the window.
5.
Replace both fuses with the new value as defined in the table
6.
Re-insert the fuse assembly.
Replacing Major Components
Changing the fuse and mains voltage
Number
Fuse type
2
250 V / 160 mA (T = slow blow)
2
115 V / 300 mA (T = slow blow)
6
6.3
above.
plug.
above.