Step 4: Check the exposure settings

Related Links
DR Detector Switch on the NX Workstation
On the NX application:
1. Check if the DR Detector Switch displays the name of the DR Detector
that's being used
2. If a wrong DR Detector is displayed, select the right DR Detector by
clicking the drop down arrow on the DR Detector Switch.
3. Check if the status of the DR Detector is ready for exposure.
On a DR Detector that has a status indicator:
Check if the status of the DR Detector is ready for exposure. If the status is
not ready for exposure, the DR Detector cannot be used for making an
exposure.
In the operator room at the X-ray generator console:
1. Check if the exposure settings displayed on the console are suitable for the
exposure.
2. If other exposure values are required than those defined in the NX exam,
use the console to overwrite the default defined exposure settings.
DX-D 300 | Operation | 55