Step 4: Check The Exposure Settings - AGFA DR 10s User Manual

Hide thumbs Also See for DR 10s:
Table of Contents

Advertisement

72 | DR 10s | Getting started

Step 4: check the exposure settings

On the DR Detector Switch:
1. Check if the DR Detector Switch displays the name of the DR Detector
that's being used
2. If a wrong DR Detector is displayed, select the right DR Detector by
clicking the drop down arrow on the DR Detector Switch.
3. Check the DR Detector Status icon.
On the X-ray system:
1. Check if the exposure settings displayed on the console are suitable for the
exposure.
2. Check if no error messages are displayed on the X-ray system.
Exposure synchronization
Depending on the configuration, the DR Detector synchronizes to the
exposure using one of these methods:
• X-ray generator synchronization
• Automatic exposure detection
Related Links
DR Detector Switch on the NX Workstation
Automatic exposure detection
0351A EN 20190319 1349 *
WARNING:
In a configuration using automatic exposure detection, the X-ray
system allows executing an exposure, even if the DR Detector is
not ready. Avoid unnecessary dose by checking the status of the
DR Detector before exposure. The DR Detector Switch displays
the DR Detector status icon.
on page 80
on page 20

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pixium 2430 ez

Table of Contents