Classification

Type of protection against electric
shock
Degree of protection against electric
shock
Degree of protection against ingress
of liquids
Methods of disinfection recommen-
ded by the manufacturer
Degree of safety of application in the
presence of flammable anesthetic
mixture with air or with oxygen or
with nitrous oxide
Mode of operation
Labelling
Remarks for HF-emission and immun-
ity
DX-D 300 | Introduction | 33
Class 1 Equipment
Type B Applied Parts
IPX0 as defined in IEC60529. Ordi-
nary equipment (enclosed equip-
ment without protection against in-
gress of liquids).
Disinfectable equipment (or ele-
ments)
Equipment for use in environments
where no flammable gases or vapors
are present
Suitable for continuous operation
• CE label: 93/42 EEC 'Medical
Devices' (Europe), EN 60601-1
• CUL label: CSA 22.2 No 601.1
(Canada)
This equipment generates, uses and
can radiate radio frequency (RF) en-
ergy and, if not installed and used in
accordance with the instructions,
may cause harmful interference to
other devices in the vicinity. In any
circumstance; however, there is no
guarantee that interference will not
occur in a particular installation.
0172E EN 20181116 1653