Scpi Commands; Calibration; Calibration:data[?] <Calibration Data Set As Definite Length Block; Calibration:data:length - R&S NRP-Z27 Operating Manual

Power sensor module
Table of Contents

Advertisement

R&S NRP-Z27/-Z37

SCPI Commands

The power sensor module R&S NRP-Z27/-Z37 is controlled via the groups of commands
CALibration (zeroing)
SENSe (measurement configurations)
SYSTem
TRIGger
SERVice.

CALibration

Table 6-1
Commands of the CALibration system
Command
CALibration
:DATA[?]
:LENGth?
:ZERO
:AUTO[?]
CALibration:DATA[?] <calibration data set as definite length block>
CALibration:DATA is used for writing a calibration data set in the flash memory of the power module
sensor.
The query yields the calibration data set currently stored in the flash memory as a definite length block.

CALibration:DATA:LENGth?

CALibration:DATA:LENGth? yields the length in bytes of the calibration data set currently stored in the
flash memory. Programs that read out the calibration data set can use this information to determine the
capacity of the buffer memory required.
1169.3870.12
Parameter
<calibration data set as definite length block>
OFF | ON | ONCE
Remote Control - Commands
Unit
Bytes
6.3
Remarks
Query only
E-2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nrp-z371169.4102.021169.3206.02

Table of Contents