Remote Control - Commands
Command
:SMOothing:STATe[?] OFF | ON
:SGAMma
:CORRection:STATe[?]
:MAGNitude[?]
:PHASe[?]
:RGAMma
:MAGNitude[?]
:PHASe[?]

SENSe:AVERage:COUNt[?] 1 to 65536

SENSe:AVERage:COUNt sets the number of measured values that have to be averaged for forming
the measurement result. The higher this averaging factor, the less the measured values fluctuate and
the longer the measurement time lasts. The parameter is rounded off to the nearest power-of-two
number.
The query yields the averaging factor used.
NOTICE
The averaging function must be activated with SENSe:AVERage:STATe ON so that the
set averaging factor becomes effective.
Default setting: 4

SENSe:AVERage:COUNt:AUTO[?] OFF | ON | ONCE

SENSe:AVERage:COUNt:AUTO activates (auto-averaging) or deactivates (fixed-filter mode) automatic
determination of the averaging factor. If auto-averaging is activated, the averaging factor is continuously
determined and set depending on the level of power and other parameters.
SENSe:AVERage:COUNt:AUTO ON activates auto-averaging and SENSe:AVERage:COUNt:AUTO
OFF deactivates it. On deactivation, the previous, automatically determined averaging factor is used in
the fixed-filter mode. The SENSe:AVERage:COUNt:AUTO ONCE command ensures that a new
averaging factor is determined by the filter automatic function under the current measurement
conditions and used in the fixed-filter mode.
The query yields
1 for OFF,
2 for ON.
Default setting: ON
1169.3870.12
Parameter
OFF | ON
0.0 to 1.0
-360.0 to 360.0
0.0 to 1.0
-360.0 to 360.0
Unit
degree
6.6
R&S NRP-Z27/-Z37
Remarks
E-2