Trigger:slope[?] Positive | Negative; Trigger:source[?] Bus | External | Hold | Immediate | Internal - R&S NRP-Z27 Operating Manual

Power sensor module
Table of Contents

Advertisement

Remote Control - Commands

TRIGger:SLOPe[?] POSitive | NEGative

TRIGger:SLOPe defines the edge of the trigger event with internal and external triggering. In this
connection, positive means increasing envelope power (with internal triggering) or increasing voltage
(with external triggering). This command has no effect in conjunction with trigger sources BUS, HOLD
and IMMediate.
The query yields
1 for POSitive,
2 for NEGative.
Default setting: POSitive

TRIGger:SOURce[?] BUS | EXTernal | HOLD | IMMediate | INTernal

TRIGger:SOURce sets the trigger source.
BUS:
Triggering with command *TRG or TRIGger:IMMediate.
EXTernal: Triggering
TRIGger:DELay and TRIGger:SLOPe.
HOLD:
Triggering only with command TRIGger:IMMediate.
IMMediate: Automatic triggering without explicit event.
INTernal:
Triggering by the measurement signal. Relevant trigger parameters:
TRIGger:LEVel, TRIGger:DELay and TRIGger:SLOPe.
The query yields
1 for BUS,
2 for EXTernal,
4 for HOLD,
8 for IMMediate,
16 for INTernal.
Default setting: IMMediate
1169.3870.12
via
USB
Adapter
6.20
R&S NRP-Z3.
Relevant
R&S NRP-Z27/-Z37
trigger
parameters:
E-2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nrp-z371169.4102.021169.3206.02

Table of Contents