R&S NRP-Z27 Operating Manual page 19

Power sensor module
Table of Contents

Advertisement

R&S NRP-Z27/-Z37
Unlatch the R&S Instrument connector on the rear panel by
Fig. 1-8
Test Setup
Fig. 1-9 shows a typical test setup:
1
R&S NRP-Z5 sensor hub
2
External power supply unit (supplied)
3
Power cable (supplied)
4
AC power supply
5
USB cable (supplied)
6
PC with USB host port
7, 8
BNC cable (optional, not supplied)
9
Trigger source (optional)
10
Triggered device (optional)
11-14
R&S NRP-Zxx power sensors (one to four sensors)
15
Signal source
16
Termination (SWR < 1.05)
The sequence in which the cables are connected is not important.
As already mentioned in section Ports and LEDs, the R&S NRP-Z4 passive interface adapter can also
be used as an alternative to a standard USB cable for connecting to the PC. This adapter is connected
to the R&S Instrument port. Simultaneous operation at two USB hosts is not possible as the USB host
connected to the R&S Instrument port always has priority.
It is possible to cascade several R&S NRP-Z5 sensor hubs by connecting the R&S Instrument port of
an R&S NRP-Z5 to one of the sensor ports of another R&S NRP-Z5. However, external triggering and
the use of the Trigger Master function are then not possible. Instead, it is recommended to connect all
R&S NRP-Z5 hubs individually to the USB host or to an interposed USB hub, and to feed the external
trigger signal to all R&S NRP-Z5 hubs via their trigger inputs; or to forward the external trigger signal
from the trigger output of the R&S NRP-Z5 to which the Trigger Master is connected to the trigger
inputs of the other R&S NRP-Z5 hubs.
1169.3870.12
pressing down the unlatching ring of the built-in
plug and pulling off the cable jack at the same
time (Fig. 1-8).
Unlatching the R&S Instrument connector
1.11
Connecting the sensor
E-6

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nrp-z371169.4102.021169.3206.02

Table of Contents