Trigger:holdoff[?] 0.0; Trigger:hysteresis[?] 0.0 To 10.0; Trigger:immediate; Trigger:level[?] X To - R&S NRP-Z27 Operating Manual

Power sensor module
Table of Contents

Advertisement

R&S NRP-Z27/-Z37
The query yields
1 for OFF,
2 for ON.
Default setting: ON
TRIGger:HOLDoff[?] 0.0 to 10.0
TRIGger:HOLDoff suppresses trigger events within the set holdoff time (in s), starting from the time of
the last successful triggering.
The query yields the set holdoff time (in s).
Default setting: 0.0 [s]

TRIGger:HYSTeresis[?] 0.0 to 10.0

TRIGger:HYSTeresis sets the hysteresis of the internal trigger threshold (parameter TRIGger:LEVel).
Hysteresis is the magnitude (in dB) by which the trigger signal level falls below the trigger threshold
(with positive trigger edge) to enable triggering again. The case is exactly the opposite with a negative
trigger edge. The trigger hysteresis setting is only relevant to the INTernal trigger source.
The query yields the trigger hysteresis in dB.
Default setting: 0.0 [dB]

TRIGger:IMMediate

TRIGger:IMMediate triggers a generic trigger event that causes the power sensor module to exit
immediately the WAIT_FOR_TRIGGER state irrespective of the trigger source and the trigger delay and
begin with the measurement. The command is the only means of starting a measurement when the
trigger source is set to HOLD.
TRIGger:LEVel[?] x to y
TRIGger:LEVel sets the trigger threshold for internal triggering derived from the test signal (in W). This
setting is irrelevant to all other trigger sources.
The query yields the trigger threshold in W.
Lower limit x and upper limit y of parameter
SENSe:CORRection:OFFSet:STATe OFF:
SENSe:CORRection:OFFSet:STATe ON:
< lower test limit >:
< upper test limit >:
Default setting: 10 × x
1169.3870.12
4.0e-6
0.4
Remote Control - Commands
x = 25 × <lower test limit>
y = <upper test limit >
x = 25 × < lower test limit > ×
10^(<SENSe:CORRection:OFFSet> / 10)
y = < upper test limit > ×
10^(<SENSe:CORRection:OFFSet> / 10)
6.19
E-2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nrp-z371169.4102.021169.3206.02

Table of Contents