Calibration:zero:auto[?] Off | On | Once - R&S NRP-Z27 Operating Manual

Power sensor module
Table of Contents

Advertisement

Remote Control - Commands

CALibration:ZERO:AUTO[?] OFF | ON | ONCE

The commands CALibration:ZERO:AUTO ON and CALibration:ZERO:AUTO ONCE perform zeroing.
For this purpose, the test signal must be deactivated or the power sensor module disconnected from
the signal source. The power sensor module automatically detects the presence of any significant
power to be measured. This causes zeroing to be aborted and error message NRPERROR_CALZERO
to be output. The CALibration:ZERO:AUTO OFF is ignored. Zeroing takes four seconds at a minimum,
but at least as long as the selected averaging filter needs for settling (only fixed-filter mode).
Repeat zeroing
during warm-up after switching on or connecting the instrument
after a substantial variation of the ambient temperature
after fastening the power sensor module to an RF connector at high temperature
after several hours of operation
when very low-power signals are to be measured, e.g. less than 10 dB above the
lower measurement limit.
For zeroing switch off the test signal and do not remove the power sensor module from
the signal source. Apart from keeping the thermal balance, this has the advantage that the
noise superimposed on the test signal (e.g. from a broadband amplifier) can be detected
on zeroing and does not impair the measurement result.
The query always yields 1 (= OFF).
Default setting
After a power-on reset, the zero offsets determined during the last calibration are used until the first
zeroing. Therefore, very slight zero offsets are to be expected with a warmed up power sensor module.
Initialization by means of *RST or SYSTem:INITialize has no influence on the current zero offsets.
1169.3870.12
6.4
R&S NRP-Z27/-Z37
E-2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nrp-z371169.4102.021169.3206.02

Table of Contents