System shut down.
High temperature of the
device
Leakage test failed.
Flow readout is always in
0 l/min.
"Device cannot be
operated"
56
Battery empty.
Ventilation slots are covered. Please ensure sufficient air
Hose system is not
completely closed.
Secretion canister is leaking.
Internal error.
Component error
Secretion has entered the
device.
Internal error.
Connect device to the supply
network. The battery is charged and
the state of charge is indicated in the
display.
ventilation.
Contact ATMOS Service or a certified
service partner. The device must be
checked.
Check the hose system and rinsing
canister for correct fit. See '3.3.2.6
Changing the secretion canister' on
page 25
Contact ATMOS Service or a certified
service partner. The device must be
checked.
1) Check whether the flow is also 0 l/
min when the system is open.
2) Contact ATMOS Service or a
certified service partner. The device
must be checked.
Contact ATMOS Service or a certified
service partner. The device must be
checked.