Display
WARNING
Device temperature too high
Provide sufficient ventilation!
Check fan!
44
Cause of error
If the device is in a tilted
position the warning „Device in
critical tilt" appears.
If the therapy has not been
started after initial operation
the warning message "Inactive
therapy" appears" without
tone.
If the therapy has already
started and has been
interrupted by the pause
symbol, the warning message
"Inactive therapy" appears with
tone.
The device must no longer be
operated.
Possible causes:
• Battery or
• pump defective.
Device temperature too high.
• Device is in the sun or near a
heater.
• Ventilation slots are covered.
• Fan is defective.
Troubleshooting
Place the device in an upright
position. The warning signal
is automatically deactivated.
Press the Play symbol to start
the therapy.
If the warning message is
suppressed, it reappears
after one minute, but without
tone.
Press the Play symbol to start
the therapy.
If the warning message is
suppressed, it appears again
with tone after one minute.
Contact the ATMOS service.
• Place the device in a cooler
location.
• Please ensure sufficient air
ventilation.
• Please contact the ATMOS
service.