Electrical Connection - Grundfos PowerAdapt Installation And Operating Instructions Manual

Table of Contents

Advertisement

Available languages

Available languages

6. Electrical connection

WARNING
Electric shock
Death or serious personal injury
Switch off the power supply before
making any electrical connections.
Make sure that the power supply can-
not be switched on unintentionally.
All electrical connections must be car-
ried out in accordance with the sepa-
rate wiring diagrams delivered with the
system.
Check and fasten the connectors with
appropriate torque when the installa-
tion is finished. The torque value is 1.2
- 1.5 Nm.
CAUTION
Hot surface
Minor or moderate personal injury
Some areas of the cabinet may get
hot. Those areas are marked with a la-
bel.
Cabinet overview of components
1
2
3
4
5
6
Pos. Description
1
Circuit breaker for MP 204
2
AC contactor
3
MP 204 control unit
4
Blocking diodes
5
DC IN/OUT
6
Terminal blocks
7
AC IN
7
Terminal blocks
1 2
3 4 5 6 7
8
Pos. Description
1
Pt100/Pt1000 sensor input
2
Pt100/Pt1000 sensor input
3
L1 output
4
L2 output
5
L3 output
6
Earth
7
Earth
8
L1
9
L2
10
L3
11
Neutral
DC input and output terminals
1
2
3
Pos. Description
1
DC+ IN
2
DC- IN
3
DC+ OUT
4
DC- OUT
9
10
11
4
9

Advertisement

Table of Contents
loading

Table of Contents