Table 3-15. Factory Configuration Messages (Continued)
Command
#CSN?
#EED 0
#EED?
#MDL str
#MDL?
#YCL
#YCL?
#YIG nrf,nrf
#YIG?
3-30
Response
#CSN nr1
Request the unit's serial number.
Default: #CSN US00000000
Example: #CSN US39440722
Load default configuration data to EEPROM.
#EED nr1
Return current value in EEPROM.
Example: #EED 43690
Set the unit's model number.
number is used to determine its VXI hexadecimal
model code. The change in model code affects the
VXI model code register when the unit is power
cycled.
Example: #MDL 'E2730A '
#MDL str, nr3
Request the unit's model number and VXI
hexidecimal model code.
Default: #MDL 'E2730A ' ,2B6
Example: #MDL 'E2730A ' ,2B6
Note: The first six (E2730A) of the seven characters
must never change. If they do, then the EEPROM
will be reset upon the next power-up of the unit.
Initiates a Calibration of the YIG DAC.
command should only be issued by qualified
personnel and only after the tuner is at normal
operational temperature.
#YCL nr1
This returns the status of the most recent YIG DAC
calibration.
Results: -1 - Complete and Uncalibrated
Example: #YCL +0
Set the YIG DAC tuning voltage values at
0 MHz and 2700 MHz respectively. This command
should only be used by qualified personnel.
#YIG nr1,nr1
Request Current YIG DAC Settings.
Example: #YIG +00000,+00000
E2730A VXI RF TUNER
Description
+0 - Complete and Calibrated
+1 - Calibration In Progress
The unit's model
This
OPERATION