Agilent Technologies E2730A Installation And Operation Manual page 52

Vxi rf tuner
Table of Contents

Advertisement

E2730A VXI RF TUNER
Command
CDE?
OPERATION
During normal operation the tuner LO status is continually monitored.
The error status bits are set to indicate the results of the monitoring
operations.
The tuner includes three command queries for determining hardware error
status. Each of the commands returns a bit mapped value between 0 and
65535, where each bit indicates the hardware status of a particular
subsystem of the tuner.
The Current Device Error (CDE) status query returns the current status of
the LO subsystems. This is not a latched register. If an LO went unlocked
and then re-locked prior to the CDE? query, there would be no evidence of
this error in the response.
The Device Dependent Error (DDE) status query returns a latched history
of any errors that have occurred since the last read of this register.
Reading the DDE clears any latched errors. If the error condition still
exists, the bit will be immediately reactivated.
The *TST command responds with same data as the DDE? query.
However, the *TST status query does not clear the latched device
dependent error status register.
The mnemonics associated with this message category are listed in
Table 3-11.
Table 3-11. Hardware Error Messages
Response
CDE nr1
Request the current device Error Register
contents. Reading this register does not affect
its contents.
where 0 = (Okay) and 1 = (Fault).
Bit
0
1
2
3
4
5
6
7
Description
This is a bit mapped response
Value
Description
1
Not Used
2
Not Used
4
Not Used
8
Not Used
16
FXE, FLEX Bootload failure
32
1LO, 1st LO unlocked
64
2LOT, 2nd LO Translation
Loop unlocked
128
2LOR, 2nd LO Resolution
Loop unlocked
3-19