Slightly wet the O-rings on the motor coupling with KAVOspray. Place the
medical device on the motor coupling and turn it until the catch audibly locks
in place.
5.2 R R emoving the medical device
Remove the medical device from the motor coupling in an axial direction.
5 5 .3 I I nserting the milling cutters or diamond grinders
Note
Only use tools that correspond to DIN EN ISO 1797-1 type 1, are made of
steel or hard metal and meet the following criteria:
- shaft diameter: 2.334 to 2.350 mm
- Overall length: max. 22 mm
- Shaft clamping length: min. 11 mm
- Blade diameter: max. 5 mm
WARNING
Use of unauthorised tools.
Injury to the patient or damage to the medical device.
▶ Observe manufacturer instructions and use the tool properly.
▶ Only use tools that do not deviate from the specified data.
DANGER
Use properly in conformance with the cutter or grinder manufacturer's in‐
structions for use.
Never use metal cutters or grinders with worn shafts can fall out during
treatment and injure the patient.
▶ Never use cutters or grinders with worn shafts.
CAUTION
Injury from using worn tools.
Tools can fall out during treatment and harm the patient.
▶ Never use tools with worn shafts.
▶ Follow the instructions for use of the tool.
CAUTION
Injury hazard from tools.
Infections or cuts.
▶ Wear gloves or fingerstalls.