Setup methods according to DIN EN ISO 17664
Setup methods according to DIN EN ISO 17664
Setup methods according to DIN EN ISO 17664
Setup methods according to DIN EN ISO 17664
S S ervicing of the clamping chuck
KaVo recommends cleaning and servicing the chuck system once a week.
Remove the cutter or grinder, place the spray nipple tip in the opening
and spray.
Carry out the servicing according to the instructions in the section,
"Care with KaVo Spray".
7 7 .5.2 C C are products and systems - Servicing: Care with KaVo SPRAYrotor
KaVo recommends servicing the product after each time it is used, i.e. after
each automatic cleaning and before each sterilisation.
Place the product on the appropriate coupling of the
KaVo SPRAYrotor and cover it with a CLEANpac bag.
Service the product.
See also: Instructions for use K K aVo SPRAYrotor
7.5.3 C C are products and systems - Servicing: Servicing with KaVo
QUATTROcare
Cleaning and care unit with expansion pressure for effective cleaning and
care.
KaVo recommends servicing the product after each time it is used, i.e. after
each automatic cleaning and before each sterilisation.
Remove cutter or grinder.
Service the product.
Chuck care
KaVo recommends cleaning and servicing the chuck system once a week.
See also: Instructions for use K K aVo QUATTROcare
Remove the cutter or grinder, place the spray nipple tip in the opening
and spray.
Subsequently treat with the care products and care systems specified.
See also: Care with KaVo QUATTROcare