12. If required, install security wire seal into and through the hole in the socket head
13. Position the wire to prevent counterclockwise rotation of the screws when the seal
14. Feed the security wire beneath the Transmitter Electronics Enclosure and through
Rosemount Gas Ultrasonic Flow Meters
screw on the Base Enclosure cover (maximum wire diameter 0.078 inch; 2.0mm).
Figure 3-34: Base Enclosure wire seal installation
A. Base Enclosure cover
B. Security wire seals (optional)
wire is taut.
the adjacent socket head screw. Twist the wire, removing all slack and seal.
Meter repairs
April 2022
89