Meter repairs
April 2022
5. Use non-fiber paper to remove any residue of acoustic coupling fluid on the surface
6. Record the serial number of the removed transducer capsule assembly and put it in
7. Repeat
3.5.2
T-200 transducer capsule assembly installation
This procedure is for installing a transducer capsule assembly into a transducer assembly
while it is installed in a meter body.
Procedure
1. Ensure that the transducer stalk, retainer and capsule assembly are clean and free of
2. Record the serial number of the transducer capsule assembly to be installed and
64
of Kapton tape.
a safe place.
Step 1
to
Step 6
if more transducer capsule assemblies are to be removed.
debris.
make sure it is correct for the intended transducer assembly.
Table 3-3: T-200 transducer stalk assembly configuration and capsule setting
Stalk assembly configuration
-01
-02
-03
-04
-05
-06
-07
-08
-09
-10
-11
-12
-13
Maintenance and Troubleshooting Manual