Emerson Rosemount 3410 Series Maintenance And Troubleshooting Manual page 33

Gas ultrasonic flow meters
Hide thumbs Also See for Rosemount 3410 Series:
Table of Contents

Advertisement

Maintenance and Troubleshooting Manual
00809-0600-3104
Table 2-3: Meter Monitor maintenance
MeterLink utility
MeterLink Logs/ Reports Menu
MeterLink Tools → Edit/
Compare Configuration menu
MeterLink Logs/ Reports Menu
Rosemount Gas Ultrasonic Flow Meters
(continued)
Diagnostics
Meter performed a Warm
start or a Warm start
required
Meter performed a Cold
start
Power failure
Troubleshooting
April 2022
Action(s)
Meter performed a Warm
start:
Collect an Archive event
log (Auditlog) using
MeterLink to view
configuration
parameter changes and
when they changed.
Warm start is required:
When you make
changes to the
transducer
characteristics, sample
rates, the device
number, or a Modbus
map file.
The meter configuration
has reset to default values
and the meter is not
configured correctly to
measure flow.
Unless the cold start
occurred after upgrading
firmware, you may need to
replace the CPU Module.
If the Cold Start occurred
after a firmware upgrade,
you must reconfigure the
meter from a previously
saved configuration file
using the Edit → Compare
Configuration screen. Then
clear the latched alarm on
the Status Summary page.
If this was a known power
fail or restart of the meter
just acknowledge this alarm
on the Status Summary
page.
If this was an unexpected
restart of the meter, verify
the integrity of the power
to the meter and make sure
that the voltage level is in
the range of 11-36 VDC at
the meter.
33