Meter repairs
April 2022
Figure 3-16: Components of the T-200 transducer assembly
A. Retaining Ring
B. Washer
C. Wave Spring
D. Transducer Housing
E. Spacer
F. Transducer Stalk Assembly
G. Transducer Retainer
3.6.1
T-200 transducer housing removal
This procedure is for removing a transducer housing from a T-200 transducer assembly
installed in a meter body.
Procedure
1. Follow
2. Hold the T-200 transducer assembly vertically with the transducer retainer facing
3. Remove the transducer retainer from the stalk. Do not drop the transducer capsule.
4. Carefully pull the transducer capsule out of the stalk and place it in a safe place.
5. Hold the transducer stalk vertically with the transducer housing facing up.
6. Place the housing removal tool on the top of the transducer housing.
7. Press the tool down while using a flat-head screw driver to release the retaining ring
8. Remove the washer and wave spring from the transducer housing.
9. Remove the transducer housing from the stalk.
10. Place the stalk upward on a flat table and do not lose the PAI spacer inside the stalk.
68
T-200 transducer assembly removal
from the meter body first. Ensure that the line pressure is down to atmospheric
pressure prior to disassembly.
up.
(see
Figure
3-16).
Note
Do not lose the retaining ring that will be popped up by the wave spring.
Maintenance and Troubleshooting Manual
to remove the transducer assembly
Emerson.com/Rosemount
00809-0600-3104