Setup 2/2; Ammeter Adjustment Setup - Agilent Technologies 4339B Service Manual

High resistance meter
Hide thumbs Also See for 4339B:
Table of Contents

Advertisement

Upon the program instruction, peform the following steps.
1. Connect the RC Box Adapter directly to the 16340A's \10 10
2. Connect the Dual Banana-BNC(m) cable to the RC Box Adapter's \Monitor H" connector.
3. Connect the other end of the Dual Banana-BNC(m) cable to the 3458A \INPUT".
Caution
To prevent electrostatic noise which may cause error in the adjustment data, do
not move the RC Box Adapter cables while the 4339B is in measurement.

Setup 2/2

This setup obtains the ammeter input resistance.
Perform the following steps as the program instructs.
1. Disconnect the Dual Banana-BNC(m) cable from the RC Box Adapter's \Monitor H"
connector.
2. Connect the Dual Banana-BNC(m) cable to the RC Box Adapter's \Monitor L" connector.
Figure 2-4. Ammeter Adjustment Setup 2/2
10 2 " connectors.
Adjustments
2-9