14. Remove the triaxial connector.
Note
When soldering the cable assembly to the triaxial connector, refer to Figure 3-3
and carefully solder the cable to the connector so that the center conductor
and the outer shield are not shorted to each other.
Figure 3-3. Cable Assembly Connection for Triaxial Connector

BNC Connector Removal

Removal Procedure

Caution
When accessing the BNC connector, use lint free gloves to avoid
contaminating the connector. Agilent Technologies supplies lint free gloves
(Agilent PN 9300-0163).
1. Remove the cover while referring to \Cover Removal".
2. Disconnect the two atcable assemblies and the wire assembly from the front side of the
A2 board.
3. Remove the screw which fastens the cable assembly on the front side of the A3 Ammeter
Board Assembly.
4. Remove the screws which fastens the cable assembly on the left front side of the chassis.
5. Remove the side trim label from both sides of the front bezel.
6. Remove the four screws that secure the front panel bezel to the chassis from both sides.
7. Place the front panel assembly, so that the front panel faces down.
8. Remove the heat shrink tube from the BNC connector being replaged.
9. Unsolder the cable from the BNC connector being replaced.
10. Unscrew the nut which connects the BNC connector to the sub panel.
11. Remove the BNC connector.