Agilent Technologies 4339B Service Manual page 23

High resistance meter
Hide thumbs Also See for 4339B:
Table of Contents

Advertisement

Perform the following steps as the program instructs.
1. Connect the 3458A \INPUT HI" post to the V Measurement Adapter \+" post using the red
lead of the Banana-Banana Jumper.
2. Connect the 3458A \INPUT LO" post to the V Measurement Adapter \Ground" post using
the black lead of the Banana-Banana Jumper.
3. Set the V Measurement Adapter's switches as follows:
4. Connect the Shunt Connector directly to the 4339B's \Ground" and \Guard" connectors.
5. Connect the V Measurement Adapter's \Output" cable to the 4339B \Output" connector.
6. Connect the V Measurement Adapter's \Interlock" cable to the 4339B \Interlock" connector.
7. Set the V Measurement Adapter's Interlock Switch \Output" to \Enable".
Table 2-2. V Measurement Adapter Switch Settings
1 2 3 4
0 1 0 1 Disable
ID
Output
Adjustments
2-7