Program Installation

The adjustment program must be installed in the computer using the following procedures.
The installation can be divided into two stages, \Making a Working Copy" and \De ning the
Con guration".

Making a Working Copy

Copy the contents of the adjustment program disk to a working disk or to the harddisk to
prevent accidental deletion or destruction of the program les. Use the working disk or the
harddisk and store the original disk in a safe place.
De ning the Con guration
De ne the con guration of the adjustment equipment. The con guration includes the GPIB
address of the Multimeter. The GPIB address will be used as a default in the program. Because
the program depends on equipment con guration, it must be rede ned if the equipment
con guration has changed.
Refer to the following procedure to de ne the con guration.
1. Turn on the computer and bring up the BASIC system.
2. Set the Mass Storage Unit Speci er to the drive/directory where the adjustment program
exists using the
MSI
3. Load the \
TE_A4339B
LOAD "TE_A4339B"
4. Run the loaded program by pressing
5. As the program instructs, enter the default GPIB address of the Multimeter.
Note
Once the default GPIB address of the Multimeter is set, it cannot be changed
unless the con guration is rede ned.

Program Execution

Perform the following steps to start the adjustment program.
Note
Install the program prior to execution. Refer to \Program Installation".
1. Turn on the computer and bring up the BASIC system. Refer to the Table 1-2 for the
required language extensions.
2. Set the Mass Storage Unit Speci er to the drive/directory where the adjustment program
exists using the
MSI
3. Load the program, \
4. Execute the loaded program by pressing
2-4
Adjustments
command.
" le from the adjustment program disk by pressing
(or
).
4
Return
5
4
Enter
5
RUN
command.
" by pressing
ADJ4339B
.
LOAD "ADJ4339B"
.
RUN
(or
).
Return
Enter
4
5
4
5

Advertisement