Ammeter Adjustment
This adjustment updates the ammeter related and voltage source related adjustment data
stored in the EEPROM. This adjustment consists of two setups.
Equipment
Setup 1/2
This setup obtains the voltage source output resistance, and the ammeter scaling factors.
Perform the following steps as the program instructs.
1. Connect the RC Box Adapter's \Input" cable to the 4339B \Input" connector.
2. Connect the RC Box Adapter's \Output" cable to the 4339B \Output" connector.
3. Connect the RC Box Adapter's \Interlock" cable to the 4339B \Interlock" connector.
4. Set the RC Box Adapter switch to \F".
2-8
Adjustments
Multimeter
RC Box
Dual Banana-BNC(m) Cable Agilent PN 11001-60001
RC Box Adapter
Figure 2-3. Ammeter Adjustment Setup 1/2
3458A
16340A
Agilent PN 04339-65005