Program Installation

The adjustment program must be installed in the computer using the following procedures.

Making a Working Copy

Copy the content of the adjustment program disk to a working disk or to the harddisk to
prevent the program le from accidental deletion or destruction. Use the working disk or
harddisk and store the original disk in a safe place.

Program Execution

Perform the following steps to start the adjustment program.
Note
Install the program prior to execution. Refer to \Program Installation".
1. Turn on the computer and bring up the BASIC system. Refer to the Table 1-2 for the
required language extensions.
2. Set the Mass Storage Unit Speci er to the drive/directory where the adjustment program
exists using the MSI command.
3. Load the program, \ ADJ4338B " by pressing LOAD "ADJ4338B"
4. Execute the program by pressing RUN .
Note
In the program, the 4338B is referred as \UUT".
Note
If the adjustment being performed follows right after EEPROM replacement,
change switch A2SW1-1 to \CLOSE", so that the 4338B will ignore the self test
errors at power-on. (Return the switch to \OPEN" when the adjustment has
been completed.) Refer to Figure B-1 for the switch location.
2-4
Adjustment
(or
).
4
Return
5
4
Enter
5