8.3 Resistance Measurement

Maximum
Range
measurement
range
400Ω
0~440Ω
3200Ω
420Ω~3600Ω
-10 ℃~ 18 ℃, +28 ℃~ 55 ℃ temperature coefficient,±0.005%FS/ ℃ .
Exciting current during measurement:
400Ω : 1.0mA±10%;
3200Ω : 0.2mA±10%;
Two-wire: Conductor resistance is excluded from errors.
Three-wire: Matching test line should be used. The total resistance of conductor should not be larger
than 25Ω.

8.4 Resistance Output

Maximum
Range
output range
400Ω
0~440Ω
3200Ω
400~3600Ω
-10 ℃~ 18 ℃, +28 ℃~ 55 ℃ temperature coefficient,±0.005%FS/ ℃ .
Resolution
2-wire,
3-wire
0.01Ω
0.25
0.1Ω
1.5
External exciting
Resolution
current
0.01Ω
0.4mA~4.0mA
0.1Ω
0.1mA~0.5mA
Accuracy (Ω)
ETX-1810
4-wire
0.15
1.0
ETX-1810
0.25
1.0
ETX-2010
2-wire,
4-wire
3-wire
0.15
0.10
1.0
0.5
Accuracy (Ω)
ETX-2010
0.15
0.50
25