6 Advanced Application

6.1 Setting 0 % and 100 % output
parameters
As for stepping operation and percentage display, 0% and 100% should be set before
using. Some gear values have been set when delivered from the factory and set values
are illustrated below:
Output Function
Resistance 400Ω
Resistance 3200Ω
Thermocouple J type
Thermocouple K type
Thermocouple T type
Thermocouple E type
Thermocouple R type
Thermocouple S type
Thermocouple B model
Thermocouple N model
The default set values may not meet your requirements, so you can reset them. Then
you can display the percentage with stepping or ramp output as well as display the
percentage. Adjust the output value with keys
for a long time until the buzzer sends a "Ticking" sound to define the new 0%
or 100% values. The new defined span value is saved in the calibrator storage
automatically and remains effective after restart. The flowing operations are available
with the setting:
DC mV
Pt100
Pt1000
Cu50
Cu100
0% value
0.00 mV
0.00 Ω
0.0 Ω
0.0 ℃
0.0 ℃
0.0 ℃
0.0 ℃
0 ℃
0 ℃
600 ℃
0.0 ℃
0.0 ℃
0.0 ℃
0.0 ℃
0.0 ℃
, press and hold
100% value
100.00 mV
400.00 Ω
3200.0 Ω
1000.0 ℃
1000.0 ℃
400.0 ℃
800.0 ℃
1500 ℃
1500 ℃
1800 ℃
1000.0 ℃
500.0 ℃
400.0 ℃
150.0 ℃
150.0 ℃
21