East Tester ETX-2710 User Manual page 18

Table of Contents

Advertisement

Figure 4.1-3
There are two kinds of cold end temperature compensation for thermocouple
measurement, automatic compensation directly using machine internal cold end
temperature compensation and manual compensation through the key to set the cold
end temperature compensation by the user.
4.1.1 Automatic compensation
First entered the thermocouple measurement mode, the default of the cold end
temperature compensation is automatic compensation reading as shown in Figure
4.1.1-1.The
sign represents that the current cold end temperature
compensation is automatic compensation. If you need further view current automatic
cold end temperature compensation value,you have to operation
key. Reading
as shown in Figure 4.1.1-1,after press
key, the
sign is replaced for
current automatic cold end temperature compensation value as
. The value
continue to display for 2s and then automatically return to the
sign.
Figure 4.1.1-1
15