  • Page 1 ETX-2710 Temperature Calibrator ETX-1710 Temperature Calibrator Users Manual Hangzhou Zhongchuang Electron Co., Ltd.
  • Page 2: Table Of Contents

    Contents 1 Basic Introduction......................1 1.1 Function....................... 1 1.2 Summary of Source and Measure Functions............1 1.3 Terminal Description....................2 1.4 Key Description....................3 1.5 Display Screen..................... 4 2 Basic Operation.......................5 2.1 Measure and Source..................... 5 2.2 Shutdown Mode....................7 2.3 Backlight Brightness Adjustment.................8 3 Function Usage.......................
  • Page 3 8.2 DC Voltage Source.....................24 8.3 Resistance Measurement..................25 8.4 Resistance Output....................25 8.5 Temperature, TC....................26 8.6 Temperature, RTD....................27 9 Product Accessories...................... 28 9.1 Standard Accessories..................28 9.2 Optional accessories...................28 10 Warning........................29...
  • Page 4: Basic Introduction

    1 Basic Introduction 1.1 Function It allows you to measure DC V, DC mV, resistance ,TC and RTD.  Outputs resistance, simulation of TC and RTD.  Thermocouple(TC) measurement provides cold-junction compensation.  Manual stepping, automatic stepping, 0~100% phase step and ramp output. ...
  • Page 5: Terminal Description

    1.3 Terminal Description Figure 1.3-1 Name Description Communication and charging Connect power adaptor to charge batteries or connect the ① connector calibrator to the computer. 3-wire and 4-wire electric and thermal resistance 3W and 4W terminal ② measurement terminal. COM Public terminal All measurement and output public terminal.
  • Page 6: Key Description

    1.4 Key Description Figure 1.4-1 Description Turns the power on or off ① Selects the measurement mode ② Selects output and analog transmitter mode. ③ Enables backlight switch display during start, enters backlight ④ brightness control mode. Selects thermal resistance mode ⑤...
  • Page 7: Display Screen

    Recovers factory default setting ⑦ Sets manual output ⑧ Cycles through: ⑨ slow repeating 0%-100%-0% ramp Fast repeating 0 % - 100 % - 0 % ramp Repeating 0 % - 100 % - 0 % ramp in 25 % steps Set output by 0% of span, Press and hold to store the source value ⑩...
  • Page 8: Basic Operation

    2 Basic Operation 2.1 Measure and Source This section acquaints you with some basic operations of ETX-2010/ ETX-1810. Proceed as follows: The connection of the calibrator as shown in Figure 2.1-1. Figure 2.1-1 Press more than 2 seconds to turn on the calibrator. The calibrator checks itself, including check on internal circuit and LCD, during which, LCD displays all contents for 1s as shown in Figure 2.1-2:...
  • Page 9 Figure 2.1-2 Then the product model(2010) and automatic shutdown time(30 min) will be displayed for 2 seconds as shown in Figure 2.1-3. Figure 2.1-3 Then the default interface of boot will be displayed as shown in Figure 2.1-4: Figure 2.1-4 Press , switch to the voltage mode.
  • Page 10: Shutdown Mode

    Figure 2.1-5 Press to increase or decrease 1 of the horizontal line position (the number will be based on automatic stepping without change of horizontal line position); press to change the position of horizontal line. Press until the buzzer works and to enter 0mV as 0% value. Likewise, press to increase output value to be 100mV, and then press until the buzzer works and enter 100mV as 100% value.
  • Page 11: Backlight Brightness Adjustment

    shutdown mode, press simultaneously. To enable the mode, press simultaneously. To adjust the time duration, press simultaneously, the screen will display as shown in Figure 2.2-1,then press and/or to adjust the time between 1 and 30 minutes and then press to store the new time duration (Without pressing any key for 5 seconds, the calibrator will quit from the adjustment automatically).
  • Page 12: Function Usage

    3 Function Usage 3.1 DC V Measurement The default function after turn on is DC V measurement. If it is required, press to re-select the voltage measurement (the display unit is V).Please connect the lines after all functions are selected. The connection mode and interface are shown in Figure 3.1-1: Figure 3.1-1...
  • Page 13: Dc Mv Measurement

    3.2 DC mV Measurement Press to re-select the DC mV function (the display unit is DC mV), the connection mode is same to that of voltage measurement. The screen will display as shown in Figure 3.2-1. Figure 3.2-1 3.3 Resistance Measurement Press to reselect the DC mV measurement function (Display unit is Ω) Figure 3.3-1...
  • Page 14: Dc Mv Output

    Figure 3.3-2 Press to force the calibrator to adopt two-wire system, three-wire system or four-wire system. Afterwards, the calibrator will not detect connecting type automatically unless you quit the mode of resistance measurement and re-enter. 3.4 DC mV Output Press to select DC mV output function, the connection mode is same to that of voltage measurement.
  • Page 15: Resistance Output

    3.5 Resistance Output Press to select the resistance output function, the connection mode is same to that of voltage measurement. Figure 3.5-1 indicates the undersize of exciting current, meanwhile, the main value flashes Press to select resistance output at 3200Ω level. Figure 3.5-2 indicates the oversize of exciting current, meanwhile, the main value flashes...
  • Page 16: Temperature Measurement

    4 Temperature Measurement 4.1 Using Thermocouple (TC) The calibrator supports eleven standard thermocouples, including J, K, T, E, R, S, B, L, U and N. Lower table summarizes the ranges and characteristics of the supported thermocouples Positive wire (H) Specified Positive lead Negative lead color...
  • Page 17 Process temperature Figure 4.1-1 Attention: Under different calibrator-thermocouple temperature environment, you shall wait for one minute to stabilize the plug temperature after connecting TC to input & output wiring. Press to select measurement mode. Press to display thermocouple reading as shown in Figure 4.1-2. If necessary, pressing continuously to select appropriate thermocouple type.
  • Page 18 Figure 4.1-3 There are two kinds of cold end temperature compensation for thermocouple measurement, automatic compensation directly using machine internal cold end temperature compensation and manual compensation through the key to set the cold end temperature compensation by the user. 4.1.1 Automatic compensation First entered the thermocouple measurement mode, the default of the cold end temperature compensation is automatic compensation reading as shown in Figure...
  • Page 19 4.1.2 Manual compensation Manual compensation through the key to set the cold end temperature compensation by the user, the following steps shall be followed: Press key to enter the set mode, reading as shown in Figure 4.1.2-1, the appear of sign shows that entering the setup mode, the assistant value means the value of manual compensation.
  • Page 20: Using Resistance Thermometer Detector (Rtd)

    4.2 Using Resistance Thermometer Detector (RTD) The calibrator accepts Pt100, Pt1000, Cu50 and Cu100, etc. The calibrator accepts RTD measurements inputs in two-, three-, or four-wire connections, with the three-wire connection the most common. A four-wire configuration provides the highest measurement precision, and two-wire provides the lowest measurement precision.
  • Page 21: Simulation Of Temperature Sensor

    Simulation temperature sensor 5.1 Simulating Thermocouples Connect calibrator input/output terminal to the instrument to be tested using thermocouple. The connecting diagram is shown below, Proceed as follows to simulate a thermocouple: Connect thermocouple to the TC input/output plughole of calibrator as shown in Figure 5.1-1.
  • Page 22: Application Of Resistance Temperature Detector (Rtd)

    then automatically return to the original display . Simulating Thermocouples also have two kinds of cold end temperature compensation, automatic and manual compensation.The using method of operation is same to thermocouple, section please refer to 4.1 for specific operation. 5.2 Application of Resistance Temperature Detector (RTD) Connect the calibrator and the instrument to be tested according to the Figure 5.2-1.
  • Page 23 Three-wire (3 W) and four-wire (4 W) terminals are just for measurement, not output simulation. The calibrator can simulate a two-wire RTD output in the front panel. To be connected to a three-wire or four-wire transducer, use folding cable to provide extra wiring.
  • Page 24: Advanced Application

    6 Advanced Application 6.1 Setting 0 % and 100 % output parameters As for stepping operation and percentage display, 0% and 100% should be set before using. Some gear values have been set when delivered from the factory and set values are illustrated below: Output Function 0% value...
  • Page 25: Auto Ramping The Output

    Press for manual steeping (increase or decrease) output based  on 25% increment. Press instantly to make output jump within the span of 0% to  100%. 6.2 Auto Ramping the Output Automatic ramping gives you the ability to continuously apply a varying stimulus from the calibrator to a transmitter, while your hands remain free to test the response of the transmitter.
  • Page 26: Power

    7 Power The calibrator needs 6 disposable LR03 model (size 7) alkaline batteries or 6 R03 model (size 7) nickel-metal hydride batteries (or nickel-cadmium batteries). The longest service life of alkaline batteries can reach 50 hours. A 12V/1A power adaptor is used for charging and providing working power for the calibrator.
  • Page 27: Specifications

    8 Specifications Specifications are based on a one year calibration cycle and apply from +18℃ to +28 ℃ unless stated otherwise. All specifications assume a 10 minute warmup period. 8.1 DC Voltage Measurement Accuracy (% of reading + Counts) Maximum Range Resolution measurement range...
  • Page 28: Resistance Measurement

    8.3 Resistance Measurement Accuracy (Ω) Maximum ETX-1810 ETX-2010 Range measurement Resolution range 2-wire, 2-wire, 4-wire 4-wire 3-wire 3-wire 400Ω 0~440Ω 0.01Ω 0.25 0.15 0.15 0.10 3200Ω 420Ω~3600Ω 0.1Ω -10 ℃~ 18 ℃, +28 ℃~ 55 ℃ temperature coefficient,±0.005%FS/ ℃ . Exciting current during measurement: 400Ω...
  • Page 29: Temperature, Tc

    8.5 Temperature, TC Accuracy (Ω) Graduation Range Resolution ETX-1810 ETX-2010 -200℃~0℃ 1.5℃ 1.0℃ 0.1℃ 0℃~1200℃ 1.0℃ 0.7℃ -200℃~0℃ 1.8℃ 1.2℃ 0.1℃ 0℃~1370℃ 1.2℃ 0.8℃ -200℃~0℃ 1.8℃ 1.2℃ 0.1℃ 0℃~400℃ 1.2℃ 0.8℃ -200℃~0℃ 1.5℃ 0.9℃ 0.1℃ 0℃~950℃ 1.0℃ 0.7℃ -20℃~0℃ 4℃ 2.5℃...
  • Page 30: Temperature, Rtd

    8.6 Temperature, RTD Accuracy (℃) ETX-1810 ETX-2010 Graduation Range Resolution 2-wire 2-wire 4-wire Output 4-wire Output 3-wire 3-wire -200℃~ Pt100 850℃ -200℃~ Pt1000 0.15 0.15 650℃ 0.1℃ Cu50 -50℃~150℃ Cu100 -50℃~150℃ 0.25 0.25 As for exciting current during measurement, please refer to resistance measurement function. As for allowable external exciting current during output, please refer to resistance output function.
  • Page 31: Product Accessories

    9 Product Accessories 9.1 Standard Accessories A set of ETX-2010/ETX-1810 calibrator also includes the following items: hard spot test leads (one set)  alligator clip (one set)  one 12V/1A power adaptor  ETX-2010/ETX-1810 users manual  Alligator Hard spot test leads Figure 9.1-1 9.2 Optional accessories 6 R03-model rechargeable batteries...
  • Page 32: Warning

    10 Warning To avoid possible electric shock or personal injury: Test a given voltage to confirm its normal operation before using.  Please follow all the safety operation standards.  Select correct function and range gear according to measurement requirements. ...
  • Page 33 Hangzhou Zhongchuang Electron Co.,Ltd. Phone: 0086-571-56770221 Fax: 0086-571-56861355 Website: Http://www.east-tester.com E-mail: sales@east-tester.com Zip Code: 310015 Address: No.3, Kangle Road, Gongshu District, Hangzhou,Zhejiang,China...