Available commands (program steps)

Heat Lid
Using
this
activate or deactivate lid heat and
the adjoining window appears.
By activating 'Lid heat on' it is
possible to enter the desired lid
temperature.
temperature must be in the range 40
- 110 °C. When Lid heat is not
activated, the lid heat is switched off.
Please note:
When reaching a 'Store' the 'Heat Lid' is deactivated automatically
Furthermore the 'Heat Lid' is deactivated automatically during temperature steps < 30
°C.
Temperature
Use this button to insert a temperature
step in the PCR program.
Temp (°C)
Set the required temperature for the
step.
Time
Set the required hold time for the step.
In hours, minutes and seconds.
Ramp-Rate
Activate the checkbox if a custom ramp rate is desired for the step. The setting here will
overwrite the 'Global Program Ramp' in the program options.
Temp-Incr./Decr
Incremental change of temperature with each successive cycle. The sign indicates whether the
temperature will increase or decrease.
Time Incr./Decr
Incremental change of hold time with each successive cycle. The sign indicates whether the
hold time will increase or decrease.