Incubation Mode
14
vwr.com I Instruction manual VWR
Run incubation
Use this button to set the device into the incubation mode. The temperature for the
thermal plate and lid can be set manually.
Setting the block and lid temperature and timer
Set temperatures of block and lid and time by entering the values of choice in the
corresponding editor fields. By entering values < 35 °C / 104 ºF the lid heating will
be turned off.
PCR Cycler XT
®
96