B.2 Measuring thermocouples
The following program measures 20 thermocouples on a TEMP 120. This example uses
CDM_TCComp()
to measure both the reference thermistor and thermocouples.
'GRANITE 6 Datalogger
'This program sets CPI Address and Device Name on a TEMP 120 with serial
'number 1234 and uses the TEMP 120 to measure 20 Type-T thermocouples.
'Declare Variables and Units
Public
TC(20) :
'Define Data Tables
DataTable(Hourly, TRUE, -1)
DataInterval(0, 60, MIN, 10)
Average(20, TC(), FP2, FALSE)
EndTable
'Main Program
BeginProg
'Set CPI Address
CPIAddModule(TEMP120, 1234, "A", 4)
Scan(5, SEC, 3, 0)
'Measure 20 Type-T thermocouples on TEMP 120 with CPI address 4
CDM_TCComp(TEMP120, 4, TC(), 20, 1,TypeT, 1, 0)
'Call Data Tables and Store Data
CallTable(Hourly)
NextScan
EndProg
CRBasic Example 3: Measuring thermocouples
Units
TC = DegC
TEMP 120